Site icon Tute Bucket

How to Apply Standard Addition Method For Spectrometric Analysis

Standard addition method of spectroscopy

Usage of standard addition method 

The standard addition method is often used in quantitative analytical chemistry. In this method, a standard is added directly to the aliquot of the analyte sample. The concentration of the analyte sample is kept constant throughout the process. Added standard concentration is varied to determine an unknown concentration of an analyte by using Beer-Lambert’s law.

Eppawala rock phosphate (ERP) is mined from Eppawala in Sri Lanka. In a dilute orthophosphate solution, ammonium molybdate reacts under acid conditions to form the heteropoly acid, molybdophosphoric acid. In the presence of vanadium, yellow vanadomolybdo-phosphoric acid is formed. This method is called the vanadomolybdo phosphoric acid method (VM method). The absorbance wavelength of the complex is in-between 460 – 480 nm. (Reference: Jeffrey G. H, Bassett J. Mendhan J.. Denney R. C.; “Vogel’s Textbook of Quantitative Chemical Analysis”, Fifth Edition (ELBS) Longman, pages 730, ISBN 0582251672).

Determination of the maximum wavelength (𝜆𝑚𝑎𝑥) for the citric-soluble phosphate

The graph of absorbance vs wavelength for the citric-soluble phosphate
Graph 01
λmax for the citric soluble phosphate  = 420 nm

Determination of the citric-soluble phosphate.

Trail Added volume of Ammonium vanadate (cm3) Added volume of Ammonium molybdate (cm3) Added volume of Eppawala Apatite (cm3) Added volume of Standard phosphate sol. (cm3) Absorbance at 420 nm wavelength
01 5.00 5.00 5.00 1.00 0.309
02 5.00 5.00 5.00 2.50 0.344
03 5.00 5.00 5.00 4.00 0.360
04 5.00 5.00 5.00 5.50 0.395
05 5.00 5.00 5.00 7.00 0.435
06 5.00 5.00 5.00 8.50 0.478
Graph 02

We can extrapolate the graph until y=0

Interpreting the Results of This Experiment

Fitting a straight line to the six absorbance readings above gives approximately A = 0.0220 V + 0.283 (R² ≈ 0.983), where V is the volume of standard phosphate solution added (cm³). Setting A = 0 gives an x-intercept of about −12.9 cm³.

This means the phosphate already present in each 5.00 cm³ portion of the Eppawala apatite solution produces the same absorbance as roughly 12.9 cm³ of the standard phosphate solution. Multiplying this volume by the concentration of the standard phosphate solution gives the amount of phosphate in the sample aliquot, which can then be converted to a concentration or a percentage of P₂O₅ in the rock.

Why Use the Standard Addition Method?

In a normal (external) calibration, standards are prepared in pure solvent and the sample is compared against them. This works well only when the sample’s other components – its matrix – do not change the signal. Real samples such as rock phosphate, soil extracts, blood or sea water often contain substances that increase or suppress the absorbance. This is called a matrix effect.

In the standard addition method, known amounts of standard are added to the sample itself. Every solution therefore has the same matrix, so any matrix effect affects the sample and the standards equally and cancels out.

Standard Addition vs External Calibration

External calibrationStandard addition
Standards prepared inPure solventThe sample itself
Corrects for matrix effectsNoYes
Sample neededSmall amountMore (one portion per addition)
Work per sampleLow – one curve for many samplesHigher – a new set of solutions for each sample
Best forSimple, clean samplesComplex samples (rocks, soils, biological fluids)

Step-by-Step Procedure

  1. Pipette the same volume of sample into a series of volumetric flasks (usually 4–6 flasks).
  2. Add increasing volumes of a standard solution (for example 0, 1, 2, 3 and 4 cm³) – one flask gets no standard.
  3. Add the same volumes of all reagents (here, ammonium vanadate and ammonium molybdate) to every flask.
  4. Dilute every flask to the same final volume with distilled water and mix well.
  5. Measure the absorbance of each solution at λmax against a reagent blank.
  6. Plot absorbance (y-axis) against the concentration or volume of standard added (x-axis) and draw the best-fit straight line.
  7. Extrapolate the line to A = 0. The magnitude of the x-intercept is the analyte content contributed by the sample.

Worked Example

10.00 cm³ portions of a water sample are placed in five 50.00 cm³ volumetric flasks. 0, 1.00, 2.00, 3.00 and 4.00 cm³ of a 10.0 ppm phosphate standard are added, the colour reagents are added, and each flask is made up to the mark.

FlaskStandard added (cm³)Added phosphate in flask (ppm)Absorbance
1000.135
21.000.200.225
32.000.400.315
43.000.600.405
54.000.800.495
  1. Added concentration in each flask = 10.0 ppm × (volume of standard ÷ 50.00 cm³), e.g. 10.0 × 1.00 ÷ 50.00 = 0.20 ppm.
  2. Best-fit line: A = 0.450 C + 0.135.
  3. x-intercept = −0.135 ÷ 0.450 = −0.300 ppm. So the sample contributes 0.300 ppm of phosphate in each 50.00 cm³ flask.
  4. Correct for dilution: the sample was diluted from 10.00 cm³ to 50.00 cm³ (×5), so the original sample contains 0.300 × 5 = 1.50 ppm phosphate.

Advantages and Limitations

Advantages

  • Cancels out matrix effects without needing to know what causes them.
  • Uses the real sample, so results are more reliable for complex materials.
  • Works with UV–Vis spectrophotometry, atomic absorption (AAS), ICP, polarography and other instrumental methods.

Limitations

  • The response must be linear over the whole range – Beer–Lambert’s law must hold.
  • Extrapolation increases uncertainty, so precise measurements matter.
  • It does not correct for a constant background signal (interference that is present even without the analyte) – a proper blank is still needed.
  • More time and more sample are needed for every unknown.

Common Mistakes

  • Not diluting all flasks to the same volume – the sample matrix must be identical in every solution.
  • Adding too little or too much standard – the added amounts should roughly span 1–3 times the expected analyte level so the line is well defined.
  • Forgetting the dilution factor when converting the x-intercept back to the original sample concentration.
  • Measuring outside the linear range of the instrument (very high absorbances, typically above about 1.0).

Frequently Asked Questions

What is the standard addition method in simple terms?
It is a calibration technique where known amounts of the analyte are added directly to portions of the sample, and the signal is extrapolated back to zero to find how much analyte was there originally.

Why do we extrapolate the graph to y = 0?
At A = 0 there would be no analyte at all. The distance from the origin to the x-intercept therefore equals the amount of analyte the sample itself contributed.

When should I use standard addition instead of a calibration curve?
Use it when the sample matrix may change the signal – for example rock, soil, food, biological or sea-water samples – or when you cannot prepare standards with a matching matrix.

Does standard addition remove all interferences?
No. It corrects for matrix effects that change the sensitivity (the slope), but not for interferences that add a constant signal. A reagent blank is still required.

Graph 03
Exit mobile version